Por favor, use este identificador para citar o enlazar este ítem: http://repositorio.ufc.br/handle/riufc/63550
Tipo: Artigo de Periódico
Título : Methodology optimization and application for measurement of gaseous elemental mercury in the amazon atmosphere
Autor : Almeida, Marcelo D.
Marins, Rozane Valente
Paraquetti, Heloísa Helena M.
Lacerda, Luiz Drude de
Palabras clave : Atmospheric mercury;Amazon atmosphere;Methodology optimization;Mercúrio - atmosfera;Amazônia - atmosfera;Metodologia - otimização
Fecha de publicación : 2008
Editorial : Journal of the Brazilian Chemical Society
Citación : ALMEIDA, Marcelo D.; MARINS, Rozane Valente; PARAQUETTI, Heloisa H. M.; LACERDA, Luiz Drude de. Methodology optimization and application for measurement of gaseous elemental mercury in the amazon atmosphere. Journal of the Brazilian Chemical Society, São Paulo, v. 19, n. 7, p. 1290-1297, 2008. Disponível em: https://doi.org/10.1590/S0103-50532008000700010. Acesso em: 18 jan. 2022.
Resumen en portugués brasileño: O objetivo desse trabalho foi otimizar uma metodologia semi-automática utilizando cartuchos comerciais preenchidos com ouro puro (Tekran Inc.) para a quantificação de mercúrio gasoso elementar em atmosfera tropical. O limite de detecção foi 0,020 ± 0,006 ng de Hg. Isso corresponde a limites de detecção variando entre 0,2 e 0,4 ng m-3, para volumes de amostragem entre 0,1 e 0,05 m3 . A similaridade entre os 9 cartuchos foi avaliada. A avaliação do efeito da vazão de amostragem entre 0,4 e 2,0 L min-1 não representou diferença. A capacidade máxima de retenção dos cartuchos (breakthrough) não foi alcançada nos experimentos com vazão de 1 ou 2 L min-1, concentração de Hg de 25 ng m-3 e volume máximo de 150 L. A condensação de água na superfície do ouro mostrou ser um problema, mas isso foi solucionado com o aquecimento do cartucho de amostragem a 60 ºC.
Abstract: The aim of this study was to optimize a semi-automatic methodology using commercial pure gold cartridges (Tekran Inc.) to quantify gaseous elemental mercury (GEM) in the Amazon atmosphere. The detection limit was 0.020 ± 0.006 ng Hg. This corresponds to detection limits varying from 0.2 to 0.4 ng Hg m-3, for field sample volumes ranging between 0.1 and 0.05 m3 . The similarity between the 9 tested cartridges was evaluated. Under sampling flow rates varying between 0.4 and 2.0 L min-1 no difference among cartridges was observed. The maximum retention capacity of the cartridge (breakthrough) was not reached in experiments with flow rates of 1 or 2 L min-1, and with Hg concentration of 25 ng m-3 and maximum air volume of 150 L. Water condensation on the gold surface showed to be a problem, but that was solved by heating the sample cartridge at 60 ºC during sampling.
URI : http://www.repositorio.ufc.br/handle/riufc/63550
ISSN : 0103 - 5053
Aparece en las colecciones: LABOMAR - Artigos publicados em revistas científicas

Ficheros en este ítem:
Fichero Descripción Tamaño Formato  
2007_art_mdalmeida.pdf383,79 kBAdobe PDFVisualizar/Abrir


Los ítems de DSpace están protegidos por copyright, con todos los derechos reservados, a menos que se indique lo contrario.